Determination of Methotrexate in human plasma by UPLC

刘盟,柳惠斌,马玉,杨建华
2013-01-01
Abstract:OBJECTIVE To establish an UPLC method to determine the plasma concentration of Methotrexate.METHODS The analytical column was Acquity UPLC BEH C18(100 mm ×2.1 mm,1.7 μm).The mobile phase was methanol-acetonitrile-the buffer of 10 mmol·L-1NaH 2 PO 4(70∶15∶15,pH6.5),and the flow rate was 0.3 mL·min-1.The detection wavelength was 306 nm,and the column temperature was 30 ℃.RESULTS The calibration curve of Methotrexate was linear within the range of 0.05-80 μmol·L-1.The absolute recovery and method recovery of Methotrexate were 82.15% and 97.30%,respectively.The intra-day and inter-day RSD were both less than 10%.CONCLUSION The method is sensitive,rapid,high efficient and reproducible,which is suitable for monitoring the blood concentration and evaluating pharmacokinetics of Methotrexate.