An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of methotrexate in human serum and plasma

Anett Engel,Lena Ruhe,Neeraj Singh,Jo Anne Wright,Franziska Liesch,Friederike Bauland,Annika I. Ostermann,Tamara Sumalowitsch,Vincent J. T. Schweinsberg,Andrea Geistanger,Johannes Kolja Hegel,Christian Geletneky,Judith Taibon
DOI: https://doi.org/10.1515/cclm-2022-1001
2023-02-15
Clinical Chemistry and Laboratory Medicine (CCLM)
Abstract:Objectives To develop an isotope dilution-liquid chromatography-tandem mass spectrometry-(ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for quantification of methotrexate in human serum and plasma. Methods Quantitative nuclear magnetic resonance (qNMR) was used to determine absolute methotrexate content in the standard. Separation was achieved on a biphenyl reversed-phase analytical column with mobile phases based on water and acetonitrile, both containing 0.1% formic acid. Sample preparation included protein precipitation in combination with high sample dilution, and method validation according to current guidelines. The following were assessed: selectivity (using analyte-spiked samples, and relevant structural-related compounds and interferences); specificity and matrix effects (via post-column infusion and comparison of human matrix vs. neat samples); precision and accuracy (in a five-day validation analysis). RMP results were compared between two independent laboratories. Measurement uncertainty was evaluated according to current guidelines. Results The RMP separated methotrexate from potentially interfering compounds and enabled measurement over a calibration range of 7.200–5,700 ng/mL (0.01584–12.54 μmol/L), with no evidence of matrix effects. All pre-defined acceptance criteria were met; intermediate precision was ≤4.3% and repeatability 1.5–2.1% for all analyte concentrations. Bias was −3.0 to 2.1% for samples within the measuring range and 0.8–4.5% for diluted samples, independent of the sample matrix. RMP results equivalence was demonstrated between two independent laboratories (Pearson correlation coefficient 0.997). Expanded measurement uncertainty of target value-assigned samples was ≤3.4%. Conclusions This ID-LC-MS/MS-based approach provides a candidate RMP for methotrexate quantification. Traceability of methotrexate standard and the LC-MS/MS platform were assured by qNMR assessment and extensive method validation.
medical laboratory technology
What problem does this paper attempt to address?