Determination of Terazosin in Human Plasma by HPLC with Fluorescence Detection

DENG Ming,ZHANG Su-fen,LIU Jian-fang,LIU Hui-chen
2012-01-01
Abstract:OBJECTIVE To establish an HPLC method with fluorescence detection for determination of terazosin in human plasma.METHODS The separation was carried out on an Apollo C18 column using the mobile phase of acetonitrile and 50 mmol·L-1 potassium dihydrogen phosphate buffer solution(22:78)at a flow rate of 1.1 mL·min-1.The column temperature was maintained at 30 ℃ and the injection volume was 10 μL.The excitation and emission wavelengths of the fluorescence detection were set at 237 nm and 370 nm,respectively.The analyte and prazosin(internal standard)were extracted with diethylether-dichlormethane(3:2) after alkalinized.RESULTS The linear range of terazosin in human plasma was 0.5-125.0 μg·L-1(r≥0.9983).The intra-and inter-day RSD were both less than 10%.The mean extraction recovery was 88.0%±4.4%.CONCLUSION The method is rapid,simple,sensitive and suitable for the clinical pharmacokinetic study of terazosin preparations.