Autoclave Sterilization of Vaseline Gauze in Primary Hospitals: An Methodology Study

董薪,刘春艳,刘则扬,张华琳,李爱国,乔军榕
DOI: https://doi.org/10.3321/j.issn:1005-4529.2003.01.019
2003-01-01
Abstract:OBJECTIVE To observe the influence of the thickness of the gauze and the location in the pot on the effect of sterilization and study the sterilization method of vaseline gauze. METHODS Put Bacillus stearother mophilas spores ATCC7953 infected fabric piece in vaseline gauze located in aluminium storage box. After steam sterilization in portable steam sterilizer, the qualitative analysis will be done. The test monitoring will be done by several times. RESULTS The result is that the spores of ATCC7953 contaminated in the gauze in layer 32 is killed completely. The killing index is (5×l0 5 5×l0 6). CONCLUSIONS Portable steam sterilizer is reliable in sterilization, however, the thickness of the vaseline gauze shall be controlled, the gauze can not be too thick. And the location of upper level in the pot is the best position for the sterilization.