Comparison of effect of two loading methods on prevention of wet pack during sterilization of paper-plastic pouch packages

Chen Mintao,Li Yuying,Zhao Hui,Sun Jiefang,Li Qiaoling
DOI: https://doi.org/10.3870/hlxzz.2012.09.007
2012-01-01
Abstract:Objective To explore methods for preventing wet pack after sterilization of metal instruments which were packaged singly with paper-plastic material and were autoclaved. Methods Totally,2 230 metal instruments packaged singly with paper-plastic material were randomized into two groups of 1 115.Packages in the control group were placed vertically and next to each other with the paper surface facing the plastic surface.For the experimental group we designed movable stainless steel frames with 2 standard widths.Packages were placed in the frame which was chosen according to specifications of instruments,with bags being spaced 2.5 cm and wider apart. Results The occurrence rates of wet pack were 36.05% in the control group and 0.54% in the experimental group,with statistical difference between the two groups(P<0.01). Conclusion Sterilization of metal instruments packaged singly with paper-plastic material by using movable stainless steel frame,can improve drying effect and prevent wet pack.
What problem does this paper attempt to address?