Sterilization efficacy of small pre-vacuum sterilizer verified by three methods

LIU Cui-mei,HU Kai,LIN Hai-yan,SHEN Shu-ming
2012-01-01
Abstract:OBJECTIVE To verify the sterilization efficacy of difference procedures in small pre-vacuum sterilizer by using chemical PCD,biological PCD and artificially inoculated dental handpiece,and select a simple,safe and feasible sterilization monitoring method.METHODS Inoculated dental handpiece,chemical PCD and biological PCD were put above the vent of the small pre-vacuum sterilizer,and the universal sterilization procedure and the rapid sterilization procedure were selected respectively.After finishing a sterilization cycle,the chemical PCD,biological PCD and dental handpiece were taken out and the results were observed and recorded.RESULTS The dental handpiece could reach SAL after sterilized in small pre-vacuum sterilizer with both sterilization procedures.The biological PCD test showed 'passed' in both procedures.The chemical PCD monitoring showed 'failed' in rapid sterilization procedure and the qualified rate was 93.3% in universal sterilization procedure.CONCLUSION Both of the universal procedure and rapid procedure could be adopted when the dental handpiece is sterilized in small pre-vacuum sterilizer.The chemical PCD monitoring is more strict and safe for sterilization.The chemical PCD monitoring can serve as an efficient monitoring system for the universal sterilization procedure.It can be adopted as sterilization monitoring method in the health facilities where can not perform the biological monitoring.
What problem does this paper attempt to address?