STUDY ON WASHING METHOD OF MEDICAL APPARATUS AND INSTRUMENT

SHAO Wei-dong,LOU Min,LI Zhen-zhen,XUE Hong-zhen
DOI: https://doi.org/10.3969/j.issn.1001-7658.2008.03.017
2008-01-01
Abstract:Objective To approach washing quality for medical instrument,and study preferred cleaning method of medical instrument.Methods To adopt test paper of occult blood for washing quality of medical instrument in spot.Results Medical instrument with soaking in enzymatic detergent plus washing of ultrasonic washing unit,eligible rate of occult blood test is 91.60%;medical instrument with soaking in hydrogen peroxide plus washing of automatic rinser,eligible rate of occult blood test is 89.47%;medical instrument with soaking in enzymatic detergent plus manual washing,eligible rate of occult blood test is 98.44%.It's showed medical instrument with dry bloodstain to amply soaking in enzymatic detergent plus washing of ultrasonic washing unit,washing quality eligible rate is 89.47%;to amply soaking in 30 g/L hydrogen peroxide plus washing of automatic rinser,washing quality eligible rate is 100%.Conclusions The washing quality is better when cleaning of contaminated medical instruments by soaking in enzymatic detergent plus ultrasonic washing unit,and good at cleaning of medical instruments with dry bloodstain by soaking in hydrogen peroxide plus automatic rinser.