Methods for Cleaning and Disinfecting of Endoscope

李六亿,巩玉秀,武迎宏,郭燕红,张泰昌,吕超英,袁洽劻,胡继红,于中麟,武沪萍,王守军
DOI: https://doi.org/10.3321/j.issn:1005-4529.2003.10.001
2003-01-01
Abstract:OBJECTIVE To try to find out the practical and reliable methods in cleaning and disinfecting endoscope in order to provide scientific basis for drafting cleaning and disinfecting rules of endoscope. METHODS To survey the disinfected efficacy using or not using enzyme of cleaning plus 2% glutaraldehyde immersion for 7 minutes in 4 hospitals and disinfected efficacy of automated washer/disinfector for endoscope in 2 hospitals in Beijing. RESULTS Qualified rates of cleaning and disinfecting were 76% and 99% in 301 cases of endoscope cleaned with enzyme and were 59% and 95% in 155 cases of endoscope cleaned with only water respectively. The difference between two groups was significant (P0.01). The qualified rate in 4 hospitals was not the same (P0.01). The qualified rate of endoscope cleaned and disinfected with machine was only 17% and much lower than that with handwork. CONCLUSIONS It is very important for endoscope to clean thoroughly and with enzyme. Gastroscope cleaned using enzyme plus 2% glutaraldehyde immersion for 7 minutes achieves standard for disinfection. Cleaned and disinfected efficacy of endoscope by handwork was better than by machine. So it should be strengthened the management of home-made automated washer/disinfector in hospitals.