Investigation of failure of flexible GI endoscope disinfection because of Burkholderia cepacia contamination in irrigation tubing

Bo Jin,Minfang Wang,Liang Huang,Xiaoyun Cheng,Xuejing Yang,Junmin Cao,Tieer Gan
DOI: https://doi.org/10.1016/j.gie.2024.05.026
2024-06-06
Abstract:Background and aims: This study aims to identify the cause of disinfection failure for multiple flexible GI endoscopes and to enhance the cleaning and disinfection procedures. Methods: Samples from the endoscopy devices, surrounding objects, cleaning water, automatic sterilizer, and integrated endoscopic washing workstation in a Digestive Endoscopy Center were collected and analyzed for microbial contamination and DNA/gene contents between May and July 2021. Results: The sample analysis revealed that the sink irrigation tubing of the washing workstation was contaminated with Burkholderia cepacia. After effective disinfection measures, the B cepacia detection in the disinfected endoscope dropped from 13.23% to 0% (P = .041). The presence of B cepacia was confirmed through homology search and gene sequencing. Conclusions: The primary reason for endoscope disinfection failure is the contamination of the sink irrigation tubing by the B cepacia bacteria. These findings emphasize the need for thorough cleaning of irrigation tubing in integrated endoscopic washing workstations, which is generally neglected in routine maintenance.
What problem does this paper attempt to address?