Monte Carlo Simulation of the Full Energy Peak Efficiency of an HPGe Detector.

Waseem Khan,Qingmin Zhang,Chaohui He,Muhammad Saleh
DOI: https://doi.org/10.1016/j.apradiso.2017.11.018
IF: 1.787
2017-01-01
Applied Radiation and Isotopes
Abstract:This paper presents a Monte Carlo method to obtain the full energy peak efficiency (FEPE) curve for a High Purity Germanium (HPGe) detector, as it is difficult and time-consuming to measure the FEPE curve experimentally. The Geant4 simulation toolkit was adopted to establish a detector model since detector specifications provided by the nominal manufacturer are usually insufficient to calculate the accurate efficiency of a detector. Several detector parameters were optimized. FEPE curves for a given HPGe detectors over the energy range of 59.50-1836 keV were obtained and showed good agreements with those measured experimentally. FEPE dependences on detector parameters and source-detector distances were investigated. A best agreement with experimental result was achieved for a certain detector geometry and source-detector distance.