HPGe Detector Efficiencies for Extended Sources Using the Monte Carlo Method

Waseem Khan,Chaohui He,Yu Cao
DOI: https://doi.org/10.1080/10420150.2019.1619733
2019-01-01
Radiation Effects and Defects in Solids
Abstract:To determine the activity of radioactive sources, accurate efficiencies are always required since the data provided by the manufacturer for detector model is insufficient to calculate the efficiencies value and obtained strong discrepancies between simulated and experimental results, especially for low photon energies. For this purpose, a detailed detector model was developed to simulate the total and full energy peak efficiencies of a p-type coaxial High Purity Germanium (HPGe) detector using GEANT4. In the simulation, gamma aqueous sources containing several radionuclides were used, covering the energy range from 59.50 to 1836.01 keV. In the case of extended sources, the coincidence summing correction factors () for two nuclides () were calculated from the peak () and total efficiencies () at 15 points throughout the source volume obtained with GEANT4. A modified detector model including the summing effect, the simulated full energy peak efficiencies agreed with the experimental values within 1%, except for the high energy range where they are less than 2% between 122 and 1836keV for the small extended source volume.