Detector efficiency calibration using 88Rb point source

C. Coşar,A. Luca
DOI: https://doi.org/10.1007/s10967-024-09478-7
2024-04-27
Journal of Radioanalytical and Nuclear Chemistry
Abstract:A coaxial type HPGe detector, and two lanthanide scintillation type detectors have been modeled and characterized by means of Monte Carlo method implemented using the MCNP code. The isotope of interest is 88 Rb which decays with energies up to 5 MeV, and can in theory aid in extending the efficiency calibration of the installations of interest in higher gamma rays energies. An 152 Eu point source was used for the experimental validation of the MCNP models and excellent agreement was observed between the experimental data and the simulated data of both 152 Eu and 88 Rb point sources. After the experimental validation of the model, the efficiency calibration was extended for three different source-to-detector distances and the results are presented in this work.
chemistry, inorganic & nuclear, analytical,nuclear science & technology