Characterization of a $^{220}$Rn source for low-energy electronic recoil calibration of the XENONnT detector

Florian Jörg,Shengchao Li,Jochen Schreiner,Hardy Simgen,Rafael F. Lang
DOI: https://doi.org/10.1088/1748-0221/18/11/P11009
2023-11-14
Abstract:Low-background liquid xenon detectors are utilized in the investigation of rare events, including dark matter and neutrinoless double beta decay. For their calibration, gaseous $^{220}$Rn can be used. After being introduced into the xenon, its progeny isotope $^{212}$Pb induces homogeneously distributed, low-energy ($<30$ keV) electronic recoil interactions. We report on the characterization of such a source for use in the XENONnT experiment. It consists of four commercially available $^{228}$Th sources with an activity of 55 kBq. These sources provide a high $^{220}$Rn emanation rate of about 8 kBq. We find no indication for the release of the long-lived $^{228}$Th above 1.7 mBq. Though an unexpected $^{222}$Rn emanation rate of about 3.6 mBq is observed, this source is still in line with the requirements for the XENONnT experiment.