Feasibility Study to use Neutron Capture for an Ultra-low Energy Nuclear-recoil Calibration in Liquid Xenon

R. Coronel,Y. Liu,M. Arthurs,S. Steinfeld,R. Gaitskell,W. Lorenzon,C.S. Amarasinghe,D.Q. Huang
DOI: https://doi.org/10.48550/arXiv.2204.03109
2022-04-06
Instrumentation and Detectors
Abstract:The feasibility of an ultra-low energy nuclear-recoil measurement in liquid xenon using neutron capture is investigated for a small (sub-kilogram) liquid xenon detector that is optimized for a high scintillation gain, and a pulsed neutron source. The measurement uses the recoil energies imparted to xenon nuclei during the de-excitation process following neutron capture, where promptly emitted $\gamma$ cascades can provide the nuclei with up to $0.3$ keV$_\text{nr}$ of recoil energy due to conservation of momentum. A successful calibration of scintillation photon and ionization electron yields below this energy will contribute to a greater sensitivity for liquid xenon experiments in searches for light WIMPs.