A 83Krm Source for Use in Low-background Liquid Xenon Time Projection Chambers

L. W. Kastens,S. Bedikian,S. B. Cahn,A. Manzur,D. N. McKinsey
DOI: https://doi.org/10.48550/arXiv.0912.2337
2009-12-11
Instrumentation and Detectors
Abstract:We report the testing of a charcoal-based Kr-83m source for use in calibrating a low background two-phase liquid xenon detector. Kr-83m atoms produced through the decay of Rb-83 are introduced into a xenon detector by flowing xenon gas past the Rb-83 source. 9.4 keV and 32.1 keV transitions from decaying 83Krm nuclei are detected through liquid xenon scintillation and ionization. The characteristics of the Kr-83m source are analyzed and shown to be appropriate for a low background liquid xenon detector. Introduction of Kr-83m allows for quick, periodic calibration of low background noble liquid detectors at low energy.