Monte Carlo Simulation of Ultra-high Resolution γ Detector

Zhao-hong MO,Zhong-hua XIONG,Zhong-guo REN,Xin-chun LAI,Qi AN
DOI: https://doi.org/10.3969/j.issn.0258-0934.2016.12.016
2016-01-01
Abstract:The ultra-high energy resolution detector based on superconducting micro-calorimeters (transition edge-sensor) has been developed for improving the NDA (non-destructive analysis) method on radiation material.The resolution is the most important parameter related to the ability of distinguish the energy peaks.Because the energy resolution of the ultra-high energy resolution detector,which is needed in detecting nuclear materials,is better than the HPGe detector,the ultra-high energy resolution detector has great promise for applications in the fields of homeland safeguard.To research on interaction between the gamma rays and the detector,this pape simulates the ultra-high resolution detector with different sources and absorbers by using MCNP5.This paper described the results of the simulation with different materials,shapes and detectors.The ultra -high resolution gamma ray detector has advantages in isotopic abundance measurement.The results can help to understand the spectral response function and to optimize the detector geometry.