Summing Coincidence Correction for Γ-Ray Measurements Using the HPGe Detector with a Low Background Shielding System

L. -C. He,L. -J. Diao,B. -H. Sun,L. -H. Zhu,J. -W. Zhao,K. Wang
DOI: https://doi.org/10.1016/j.nima.2017.09.043
IF: 1.335
2017-01-01
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Abstract:A Monte Carlo method based on the GEANT4 toolkit has been developed to correct the full-energy peak (FEP) efficiencies of a high purity germanium (HPGe) detector equipped with a low background shielding system, and moreover evaluated using summing peaks in a numerical way. It is found that the FEP efficiencies of Co-60, Ba-133 and Eu-152 can be improved up to 18% by taking the calculated true summing coincidence factors (TSCFs) correction into account. Counts of summing coincidence gamma peaks in the spectrum of Eu-152 can be well reproduced using the corrected efficiency curve within an accuracy of 3%. (c) 2017 Elsevier B.V. All rights reserved.