Preliminary Performance Evaluation of ECAT HR+ PET

管一晖,林祥通,赵军,陆汉魁,王红兵,左传涛,刘永昌
DOI: https://doi.org/10.3760/cma.j.issn.2095-2848.2001.05.021
2001-01-01
Abstract:Objective To evaluate the performance of the high resolution whole body PET scanner ECAT EXACT HR +. Methods The scanner consists of four rings of 72 bismuth germanate block detectors each, covering an axial field of view of 15.5 cm with a patient port of 56.2 cm. A single block detector is divided into an 8×8 matrix, giving a total of 32 rings with 576 detectors each. The scanner is equipped three 68 Ga line sources for transmission scans and daily quality control. The spatial resolution, scanner fraction, count rate, sensitivity and uniformity were evaluated after the National Electrical Manufacturers Association (NEMA) protocol using the standard acquisition parameters. Results The transaxial resolution in the two dimensional mode was 4.50 mm in the center and increased to 4.63 mm at a distance of r=10 cm from the center. The axial slice width measured in the two dimensional mode was 4.05 mm FWHM. In the three dimensional mode the average axial resolution varied between 4.53 mm FWHM in the center and 4.76 mm at r=10 cm. The scatter fraction was 15.1% for a lower energy discriminator level of 350 keV. The total system sensitivity for true events was 6.167+05 MBq·mL -1 ·min -1 for 2D. From the uniformity measurements, we obtained a volume variance of 3.6% for 2D. Conclusions The ECAT EXACT HR + has an excellent, nearly isotropic spatial resolution, which is advantageous for brain and even small animal study. The scanner offers a sufficient sensitivity and count rate capacity for three dimensional whole body imaging. Compared with the performance data provided by the CTI the data of our PET scanner are similar to the CTI data.
What problem does this paper attempt to address?