Performance Evaluation of A New High-Sensitivity Tof Clinical Pet/Ct System

Li Huo,Ruixue Cui,Haiqun Xing,Nan Li,Cheng Zhu,Heyu Wu,Yuan-chuan Tai,Fang Li,Hui Zhang
2015-01-01
Journal of Nuclear Medicine
Abstract:433 Objectives PoleStar m660 is a newly developed high sensitivity PET/CT system with TOF capability. The aim of this study is to characterize the performance of the new system. Methods The PoleStar m660 scanner combines a 64-slice CT with a whole-body PET. The PET subsystem consists of 24 detector buckets, each is organized as 2 detector modules along the trans-axial direction and 3 or 4 detector modules along the axial direction. (A 4-ring configuration was evaluated in this study.) Each detector module contains a 14×14 LYSO array, with a crystal size of 3.63mm×3.63mm×20mm. Performances of the PET subsystem were measured using the NEMA NU 2-2007 standards. A 425-650 keV energy window and a 4.1 ns timing window were used. The timing resolution was measured using a rotating line source that orbits around the center of field-of-view (CFOV) at a radius of 20 cm. Coincidence events from all lines-of-response that pass the CFOV were sorted to create the timing spectrum and to estimate the system timing resolution. Results The transverse (axial) spatial resolutions were 3.76 (3.64) mm and 4.56 (5.29) mm FWHM at 1 cm and 10 cm off the CFOV, respectively. The measured sensitivity was 10.9 kcps/MBq at the CFOV and 10.7 kcps/MBq at 10cm off the CFOV. The peak NECR was 224.6kcps at an activity concentration of 29.0 kBq/ml. For the image quality phantom, the contrast recovery ratios ranged from 70.2% to 94.3%, while the background variability ranged from 7.8% to 2.1%. For the timing resolution, an average of 434 ps FWHM was measured. Conclusions The physical performances of the PoleStar m660 PET/CT system were characterized. Results showed improved system sensitivity, count rate performance, and timing resolution when compared with currently available commercial clinical PET systems. Clinical evaluation of the scanner is currently underway and the results will be presented.
What problem does this paper attempt to address?