Research on Performances of PET/CT with Time-of-flight Capability

Yan Zhang
2012-01-01
Abstract:Objective:The performances of a PET/CT scanner with time-of-flight(TOF) capability were measured and studied.Methods:Performances measurements on the scanner were made using the National Electrical Manufactures Association(NEMA) NU22007 procedures and Company's Performance Test Manual.The energy/timing resolution,spatial resolution,sensitivity,the scatter fraction,counts performance and imaging quality were measured and analyzed.Different types of PET/CT were compared and the imagingcapabilities of TOF were discussed.Results:The system energy and timing resolution were measured to be 11.59% and 499.3 ps.The transverse and axial resolutions near the center were 4.54mm and 4.85mm,whereas at 10 cm were 5.13 mm,5.55 mm,5.76 mm.The absolute sensitivity of this scanner was 6.4(center) and 6.8(10 cm) cps/kBq,whereas scatter fraction was 32%.The peak NECR was 97.6kcps and occured at 14.03kBq/mL.Image quality measurements with phantom showed that the 10 mm diameter lesion could be clearly distinguished.Conclusion:The PET/CT which achieved TOF capability performed good characteristic,especially in the improvement of image quality.And shorter scan times and lower dose could be gained at the same image quality.