Study of Time-of-flight Influence on Image Quality by PET/CT Scanner

WANG Feng,ZHANG Yan,ZHAO Wei,ZHU Hua,XIE Qing,ZHOU Ni-na,YU Jiang-yuan,LI Nan,WANG Xue-juan,YANG Zhi
DOI: https://doi.org/10.7687/J.ISSN.1003-8868.2013.01.008
2013-01-01
Abstract:Objective To investigate the benefits of time-of-flight(TOF) technology on image quality in a PET/CT scanner,and to apply Monte Carlo method to simulating some performances of PET/CT with TOF capability.Methods Philips GEMINI TF PET/CT was used for phantom scan studies.The raw data were acquired and reconstructed using TOF and non-TOF methods.The contrast of TOF and non-TOF images were compared and analyzed in different experiment conditions.The PET/CT scanner's geometry and physics models were established using Monte Carlo method and the PET/CT counting performance and signal-to-noise(SNR) were simulated in different time resolutions.Results Research results stated that PET images quality was clear and contrasts were good with TOF technology.The contrast could be improved by 10%,especially in low counts and small lesion.Monte Carlo simulation showed shat coincidence counts and SNR increased with activity.However,in higher activity region there were much more random counts and the SNR increased slowly.The relation between SNR and the time resolution was that the less the time resolution,the higher the SNR gained.The SNR of 520 ps time resolution was improved by 50% larger than that with 1 000 ps time resolution.Conclusion The PET/CT image quality can be improved using TOF technology,and short scan time and low dose can be gained at the same image quality.