Simulation Study of a Non-Human Primate PET System Based on a DOI-Capability Detector Block

Li Cheng,Mengle Xue,Yingcai Ji,Shuai Huang,Nianming Jiang,Yaqiang Liu
DOI: https://doi.org/10.1109/nss/mic44867.2021.9875861
2021-01-01
Abstract:In this work, a half-meter-long PET scanner was designed based on a 7mm-DOI-capability detector block to achieve both high spatial resolution and high sensitivity for non-human primate (NHP) imaging. The scanner consists of 15 block rings in axial and 36 blocks per ring. The performance of the proposed design was evaluated based on GATE simulations. The sensitivity at the center of field-of-view (FOV) was 19.3% with a 250~750 keV energy window. A mini-Derenzo phantom study demonstrated that 1.0 mm hot rods could be recovered at the FOV center, while 1.4 mm hot rods could be partially identified near the edge of 30 cm transverse FOV. NCAT phantom was zoomed out and simulated to evaluate the imaging performance of short frame time. Besides, we compared the proposed design with the mini-EXPLORER II PET scanner by calculating Cramer-Rao lower bound. The result demonstrated that although lack of TOF capability, the proposed design performed better at larger radial offset (>~6 cm) in the transverse plane, and it also outperformed in the axial direction because it suffered less from axial parallax error and the compact design led to higher sensitivity.