Performance Evaluation Of The Hamamatsu Shr22000 Whole-Body Pet Scanner Using The Iec Standard

Tian Yi,Katabe Atsushi,Liu Hua-Feng,Bao Chao,Yamashita Takaji
DOI: https://doi.org/10.3321/j.issn:0254-3052.2006.11.019
2006-01-01
Abstract:SHR22000 is a whole-body PET scanner developed by Hamamatsu Photonics K.K. The scanner has a 600mm patient aperture and a 225mm axial FOV, which can, cover the whole human head and large portion of the body so that a common whole-body scan can be accomplished within 5 bed position with a 3.2mm average transverse resolution at 10mm from the axis of FOV. The paper gives a system performance evaluation of this scanner using GB/T 18988.1-2003/IEC 61675-1:1998 standard, which includes the spatial resolution, tomographic sensitivity, count rate characteristic, scatter fraction and recovery coefficient. At last a human body scanning result is given as an imaging example.