Performance Evaluation of Stationary Dedicated Cardiac SPECT with Multi-Pinhole Collimators: Comparison with Conventional SPECT and CZT SPECT

Haipeng Wang,Feng Guo,Xinqiang Wang,Zongyao Zhang,Daoyu Wang,Yingying Xu,Yaqiang Liu,Si Chen
2018-01-01
Journal of Nuclear Medicine
Abstract:513 Objectives: The conventional low-energy high-resolution (LEHR) parallel-hole SPECT for nuclear cardiac imaging suffered from low sensitivity and spatial resolution. We designed and implemented a high sensitivity dedicated cardiac multi-pinhole collimator, which is capability for dynamic studies, based on the clinical variable-angle dual-head SPECT ImageE NET 632. The purpose of our study is to evaluate the performance of this multi-pinhole cardiac SPECT system (MP-SPECT). Methods Two SPECT detector heads were placed in 76 degree setup. Conventional parallel-hole collimators were replaced with the multi-pinhole collimators. Each collimator had twelve 4.5 mm tungsten pinholes and all the pinholes focused on the heart region without projection overlap. Five separate and parallel line sources phantom, injected by 2 mCi 99mTc, was scanned for 5 min and reconstructed with 80 MLEM iterations. The average FWHM in transversal image for each line was calculated and used to evaluate the spatial resolution. The volume sensitivity per axial centimeter (VSAC) was measured according to NEMA NU1-2012 with a scaled phantom. A heart phantom with two myocardium defects (2 cm and 1.5 cm in length), injected by 2 mCi 99mTc, was scanned for 5 min and reconstructed with 80 iterations. Two patients with suspicious of coronary heart disease were carried out MP-SPECT (5min) and reconstructed with 80 MLEM iterations. As compared, the patients also underwent MPS imaging on LEHR-SPECT BrightView XCT (15min), D-SPECT (5min) and Discovery NM 530c (5min) according to standard 1-day stress protocol. The reconstruction images were compared. Results The reconstruction spatial resolution (average FWHM of line sources) is 6.0 mm, better than LEHR SPECT (~14 mm). The measured VSAC is 92 s-1/(kBq[asterisk]cm2), about four times higher than LEHR SPECT typically 20 s-1/(kBq[asterisk]cm2 ). Reconstruction image of heart phantom was shown in Figure 1. The shape of myocardium was well reconstructed and both the two defects could be identified clearly. The results of patients study were shown in Figure 2 and Figure 3. Visually, the spatial resolution of MP-SPECT was better than LEHR-SPECT (Figure 2) and comparable to CZE SPECT (Figure 2). Conclusions The proposed MP-SPECT system has higher sensitivity and better spatial resolution than conventional LEHR-SPECT. The myocardial perfusion scans can be carried out without rotation and within 5 min. The heart phantom study show good defect identification and the patients study show that the image quality of MP-SPECT is better than LEHR-SPECT and comparable to commercial dedicated cardiac CZT SPECT.
What problem does this paper attempt to address?