Research on verifications of physical parameters of virtual wedge on Siemens accelerator by Mapcheck

Jun LU,Xiu-chun ZHANG,Jun-xin WU,Su-fang QIU,Jian-ji PAN
DOI: https://doi.org/10.3969/j.issn.1673-5269.2008.14.017
2008-01-01
Abstract:Objective: To verify the veracity of the arithmetic model of virtual wedge of three-dimensional treatment plan system (3D TPS) ADAC Pinnacle (V7. 4f) by a 2D diode array detector Sun Nuclear Mapcheck( Model 1175). Methods: By using a 2D diode array detector Sun Nuclear Mapcheck (Model 1175), the 1 vw and 2 vw direction virtual wedge field of 15°, 30°, 45° and 60° virtual wedge of Siemens Primus 6 MV and 18 MV, SSD 100 cm, depth 5 cm and 10 cm, field size 15 cm × 15 cm were measured, and then compareed with the same virtual wedge field produced by three-dimensional treatment plan system ADAC Pinnacle (V7. 4f). Results: The virtual wedge was almost equal to 1, with a maximal variation of 0.0119 no matter what the value of beam energy, field size or wedge angle was. The profile of wedge direction had a good conformation between those measured by Mapcheck and produced by 3D TPS with a maximal variation of - 3.8%. Conclusions: There is a good conformation between those parameters metioned above measured by Mapcheck and produced by 3D TPS, and the arithmetic model of virtual wedge of ADAC Pinnacle (V7. 4f) is permited for the application of clinic. The 2D diode array detector Sun Nuclear Mapcheck (Model 1175) is a good implement for daily QA.
What problem does this paper attempt to address?