A New Flow Injection Calibration System Based On Automatic Diagnosis And Correction For Fault Peak Signal Using Flow Injection Gradient Information

Sh Fan,Zl Fang
DOI: https://doi.org/10.3321/j.issn:0253-3820.2002.09.006
IF: 1.193
2002-01-01
Chinese Journal of Analytical Chemistry
Abstract:A new flow injection (FI) calibration method was proposed for spectrophotometry based on exploitation of gradient information in FI peak profile, which involved on - line automatic diagnosis and correction for faulty peak signal using a reference peak. Calibration curves were constructed by using the mean of ratios of absorbances contained in standard peak profiles versus those of reference peak. The precision for present method is equivalent to that of standard series calibration. The calibration method seems to be more suitable for some industrial process monitoring, due, to the capabilities of on-line diagnosis and correction automatically for faulty peak and simultaneously utilizing the remained information contained in the faulty peak.
What problem does this paper attempt to address?