Flow-injection calibration procedures with the use of fully rotary valve

Pawe? Ko?cielniak,Joanna Janiszewska,Zhaolun Fang
1996-01-01
Abstract:An injection valve for flow injection analysis has been developed that is fully rotary, i.e. a rotor is capable of rotating full circle in relation to the stator. The valve has been applied for improving the calibration procedures. Instead of using as usually a set of standard solutions a single standard solution was prepared and injected from loops with different volumes. By doing so a set of signals was produced enabling to draw a calibration graph. The performance of the valve has been presented in two different modes adapted to execute both the calibration curve method and the standard addition method. The calibration procedures appeared to be comparable with those conventionally performed in terms of accuracy and precision but better regarding consumption of time, work and reagents.
What problem does this paper attempt to address?