The Application Of Psd In The Collimation Testing

Z Zhang,Lh Jiao,Xd Hu
DOI: https://doi.org/10.1117/12.293491
1997-01-01
Abstract:In this paper a detecting system that detects the errors of the dynamic angle measuring instrument has been developed. Theoretical analyses have been done on the detecting pinciples of the system. This new automatic testing system in which PSD insteads of collimator can measure the dividing errors of the dynamic angular instrument.This detecting device overcomes the shortage of the collimator whose dynamic respond spectrum is limited in the couse of dynamic testing. Moreover, this device has simple structure and is easy to adjust. The system consists of two - mode laser, a laser collimation system, rotating table, PSD and a 12 - sides polygon mirror that was used as an angle standard.The opto-electronic Position - Sensitive Detector(PSD) has the advantage of short respond time, high discrimination and large respond spectrum. But it has large position testing error. In this paper, the position testing error of PSD has got demarcating correction in order to meet the need of high precise measurement.This device can work in high speed. Its accuracy is high in the dynamic testing. With the system, a preliminary experimental research has been made on the round synchronous transformer. Experimental results show that the system satisfies the initial design requirements.