Research on the measuring procedure of dual energy X-ray absorptiometry (DXA) in determining the bone mineral density (BMD)

Linlin Qin,Jingbiao Chen,Wei Zhang,Haibo Ma,Chunhua Ge,Yexia Xiao
1996-01-01
Abstract:The influence of measuring procedure on the determination precision of BMD was studied, and it was found that the position of measured subjects and the analysis of acquired images are two main influential factors. For example, the precision of BMD values obtained from images analysed by two technicians is lower than that by one technician; for DXA of Lunar DPX-L, the vitro precision is better than 1%, and the vivo precision expressed by coefficient of variation (CV) are 0.75% at lumbar, 0.47% at neek, 1.14% at Ward's triangle, and 0.72% at trochanter respectively. hence it was proposed to establish the standard phantom of our country.