The Protocol of Ultrasonic Backscatter Measurements of Musculoskeletal Properties

Dongsheng Bi,Lingwei Shi,Boyi Li,Ying Li,Chengcheng Liu,Lawrence H. Le,Jingchun Luo,Sijia Wang,Dean Ta
DOI: https://doi.org/10.1007/s43657-023-00122-0
2023-01-01
Phenomics
Abstract:This study aims to introduce the protocol for ultrasonic backscatter measurements of musculoskeletal properties based on a novel ultrasonic backscatter bone diagnostic (UBBD) instrument. Dual-energy X-ray absorptiometry (DXA) can be adopted to measure bone mineral density (BMD) in the hip, spine, legs and the whole body. The muscle and fat mass in the legs and the whole body can be also calculated by DXA body composition analysis. Based on the proposed protocol for backscatter measurements by UBBD, ultrasonic backscatter signals can be measured in vivo, deriving three backscatter parameters [apparent integral backscatter (AIB), backscatter signal peak amplitude (BSP A ) and the corresponding arrival time (BSP T )]. AIB may provide important diagnostic information about bone properties. BSP A and BSP T may be important indicators of muscle and fat properties. The standardized backscatter measurement protocol of the UBBD instrument may have the potential to evaluate musculoskeletal characteristics, providing help for promoting the application of the backscatter technique in the clinical diagnosis of musculoskeletal disorders (MSDs), such as osteoporosis and muscular atrophy.