Ultrasonic Backscatter Theory, Method, and Diagnostic Instrument for Cancellous Bone

Dean Ta,Chengcheng Liu
DOI: https://doi.org/10.1121/1.4970037
2016-01-01
Abstract:Recently, significant progress has been made in the backscatter measurement of cancellous bone. In this work, we will introduce the basic theory and some key techniques for backscatter cancellous bone evaluation. Backscatter measurements in vivo were also carried out with a diagnostic instrument, and the feasibility of using the backscatter for cancellous bone evaluation was investigated. Researchers had reported conflicting associations between backscatter and bone features. Our studies demonstrated that this phenomenon was due to different signal selections. The dense cortical shell obstructs the ultrasonic evaluation of the underlaid cancellous bone in vivo. We determined the influence of the overlying cortex on the backscatter signal and proposed a compensation method for the cortical influence. Using an ultrasonic backscatter bone diagnostic instrument, the measurements were performed on 1226 adults and on 467 neonates in the calcaneus. For the adults, the backscatter was significantly correlated with the bone mineral density (BMD) of the spine and hip. The backscatter parameters were moderately correlated with the gestational age and birthweight in the neonates. The in vivo findings demonstrate the ultrasonic backscatter as a useful adjunct measurement for the multi-feature assessment of cancellous bone.
What problem does this paper attempt to address?