Ultrasonic Backscatter Measurements of Human Cortical and Trabecular Bone Densities in a Head-Down Bed-Rest Study.

Dongsheng Bi,Zhongquan Dai,Duwei Liu,Feng Wu,Chengcheng Liu,Ying Li,Boyi Li,Zhili Li,Yinghui Li,Dean Ta
DOI: https://doi.org/10.1016/j.ultrasmedbio.2021.04.002
IF: 3.694
2021-01-01
Ultrasound in Medicine & Biology
Abstract:This study aims to investigate the feasibility of quantitative ultrasonic backscatter in evaluating human cortical and trabecular bone densities in vivo based on a head-down-tilt bed rest study, with 36 participants tested through 90 d of bed rest and 180 d of recovery. Backscatter measurements were performed using an ultrasonic backscatter bone diagnostic instrument. Backscatter parameters were calculated with a dynamic signal-of-interest method, which was proposed to ensure the same ultrasonic interrogated volume in cortical and trabecular bones. The backscatter parameters exhibited significant correlations with site-matched bone densities provided by high-resolution peripheral quantitative computed tomography (0.33 < |R| < 0.72, p < 0.05). Some bone densities and backscatter parameters exhibited significant changes after the 90-d bed rest. The proposed method can be used to characterize bone densities, and the portable ultrasonic backscatter bone diagnostic device might be used to non-invasively reveal mean bone loss (across a group of people) after long-term bed rest and microgravity conditions of spaceflight missions.