A Sub-Millimeter Spatial Resolution Achieved by a Large Sized Glass RPC

Li Qi-Te,Ye Yan-Lin,Ji Wei,Wen Chao,Liu Hong-Tao,Ge Yu-Cheng
DOI: https://doi.org/10.1088/1674-1137/37/1/016002
IF: 2.944
2013-01-01
Chinese Physics C
Abstract:Three large sized glass resistive plate chambers (RPCs) are built and applied to measure the spatial resolution of the detector. The readout strips are collected to a LC delay-line and the time difference is used to determine the position. Cosmic rays are triggered by a set of two scintillation counters and the coincidently measured positions from the three RPCs are used to deduce the position uncertainty. In average a spatial resolution of 0.90 mm (FWHM) is obtained for a single RPC, with a good uniformity across the detection area. This result suggests that large sized glass RPC operating in the avalanche mode is a promising candidate for the muon tomography detection system.
What problem does this paper attempt to address?