A Large Area MRPC Designed for Muon Tomography

WANG Jie,FENG Sheng-qin,WANG Yi,WANG Xue-wu,ZENG Ming,XIE Bo
DOI: https://doi.org/10.3969/j.issn.0258-0934.2016.02.014
2016-01-01
Abstract:We have developed a large area two -dimensional high position resolution MRPC for muon tomo-graphy .The sensitive area of the module is 1 072 mm ×1 072 mm, through a cosmic-ray experiment , the per-formance of MRPC module is carefully observed and each channel is calibrated .Through an X ray experiment with a narrow slit , the position resolution is studied .The results show that the efficiency is higher than 98%and the spatial resolution can reach 272 μm.