Determination Partition Coefficients of Xenon in Lipid Emulsion of Different Concentration by GC-MS

CAO Wei,LI Zhang-wan,LIU Jin,ZHANG Wen-sheng
DOI: https://doi.org/10.3969/j.issn.1006-0103.2007.04.023
2007-01-01
Abstract:OBJECTIVE To establish a GC-MS method for determination of partition coefficients of Xe in lipid emulsion.METHODS HP-5MS capillary column(30 m×0.32 mm×0.5 μm)was used,with column temperature of 40℃,injector temperature at 120℃,split ratio of 100:1,MS as the detector,high purity helium(1.5 ml·min-1)as carrying gas.MS conditions were ionization mode EI,the source temperature at 230℃,electron energy of 70 eV and SIM(m/z)of 129 and 132.The partition coefficients of xenon in 10%,20% and 30% lipid emulsion were determined by syringe-syringe two-stage headspace equilibration technique.RESULT The calibration curve of xenon had good linear relationship in the range of 3.125%-0.012%(r=0.998).The limit of detection was 2×10-6.RESULTS The partition coefficients of xenon in 10%,20% and 30% lipid emulsion were proportional to the content of the soybean oil in lipid emulsion.CONCLUSION As the carrier of xenon,the possibility of using lipid emulsion to produce xenon vein preparation is low.If xenon should be delivered by intravenous injection, a better carrier system that has the higher xenon solubility should be searched out for.
What problem does this paper attempt to address?