SIMULTANEOUS AND QUICK DETERMINATION OF N-HEXANE,BENZENE SERIES,ESTER AND KETONE IN AIR BY FILLING COLUMN CHROMATOGRAPHY
朱惠芳,胡美华,李良,江勇
2010-01-01
Abstract:[Objective] To develop a method for the simultaneous determination of n-hexnae, acetone, butanone, ben-zene, 4-methyl-2-pentanone, toluene, butyl acetate, m-xylene and o-xylene in air by stainless filling gas chromatography. [Methods] N-hexnae, acetone, butanone, benzene, 4-methyl-2-pentanone, toluene, butyl acetate, m-xylene and o-xylene in air were collected with charcoal tube, then were injected after being desorbed with carbon disulfide, through stainless column filling FFAP: 6201 red carrier that was 5︰100 (2m ×4mm). The primary temperature set at 40℃ remaining 1.5 minutes, and the middle temperature was 50℃ remaining 3 minutes with elevate temperature ratio 20℃ per minute, and the last temperature was 100℃ remaining 2 minutes with elevated temperature ratio 10℃ per minute. [Results] Retention time for nhexnae, acetone, butanone, benzene, 4-methyl-2-pentanone, toluene, butyl acetate, m-xylene and o-xylene were 0.868, 2.00, 2.88, 4.95, 6.81, 3.39, 5.76, 8.21 and 9.08 min, respectively. The linear ranges were 8.33-266.7, 8.44- 270.0, 7.50-240.0, 9.38-300.0, 9.58-306.7, 7.70-246.4, 7.42-237.6, 8.92-285.6 μg / ml and 5.98-191.2 μg / ml. The lim- its of detection were 0.3, 6.8, 4.0, 1.0, 0.5, 0.7, 1.2, 2.5 and 2.5 μg / ml. The coefficients of correlation were 0.999 3, 0.999 2, 0.999 3, 0.999 8, 0.999 8, 0.999 8, 0.999 8, 0.999 8 and 0.999 8. [Conclusion] The method is applicable in simultaneously determining n-hexnae, acetone, butanone, benzene, 4-methyl-2-pentanone, toluene, butyl acetate, m-xylene and o-xylene in air.