Determination of cineole in ligustrazine transdermal patches by gas chromatography

Jian-ming HUANG,Hui-nan XU,Wei-yu WENG,Teng SHEN,Yu FANG,Bing WANG
DOI: https://doi.org/10.3969/j.issn.1007-7669.2005.02.004
2005-01-01
Abstract:AIM: To develop a method for the determination of cineole in ligustrazine transdermal patches. METHODS: The samples were extracted ultrasonically with methanol and were analyzed by gas chromatography. The chromatography separation was performed on a HP-50 capillary column (30 m×0.25 mm, 0.25 μm) using nitrogen as a carrier gas with the constant flow-rate of 0.4 mL·min -1. The column temperature was set at 30 °C initially and then ramped to 120 °C at the rise rate of 2 °C·min -1. The injector temperature was 250 °C and the temperature of flame ionization detector (FID) was 280 °C. RESULTS: Calibration curve was linear with r= 0.999 2. The recovery of the method ranged from 96.3 % to 107.8 % and RSD 2.53 %. CONCLUSION: The method proved simple, specific and applicable to the determination of cineole in ligustrazine transdermal patches.