Determination of Residual Solvents in Lamivudine by Headspace Gas Chromatography

倪珍珍,王翰华,姚彤炜
DOI: https://doi.org/10.3969/j.issn.1007-7693.2008.04.022
2008-01-01
Abstract:OBJECTIVE To develop a method for detecting the residual solvents in lamivudine.METHODS The determination was performed on HP-5 capillary column(5% phenyl methyl siloxane,30 m×320 μm),the column was subjected to temperature programming,the detector temperature was 200 ℃,the carrier gas was nitrogen,the flow rate was 7.8 mL·min-1 and the split ratio was 5∶1. RESULTS All the four residual solvents were well separated in 8 minutes,the intra-day variability were 8.3%,the limits of detection were 0.4~4.0 μg·mL-1.The sample met the requirements of this test.CONCLUSION The method can be used to determine the residual solvents in lamivudine.