Determination of residual organic solvents and solvent impurities in rhein by capillary gas chromatography with headspace sampling

WU Xiu-lan,WANG Hao,LIU Zhi-hong,ZHANG Hai-tao,DU Ying-xiang,YE Wen-cai
DOI: https://doi.org/10.3969/j.issn.1006-3765.2008.05.013
2008-01-01
Abstract:OBJECTIVE To establish a headspace capillary gas chromatography method for quantification of residual organic solvents and solvent impurities(chloroform,dichloromethane,1,2-dichloroethane,carbon tetrachloride)in rhein.METHODS The determination was performed on a 5% phenyl methyl siloxane column(HP-5,30 m×0.32 mm,0.25 μm)with a constant column temperature at 40 ℃ and a constant column pressure of 88 kPa.Nitrogen was chosen as the carrier gas.The injection port temperature was 200 ℃.The ECD detector worked at the temperature of 300 ℃.The headspace injection was applied with the injection volume of 1mL and the split ratio was set at 9∶1.NaOH solution(0.05 mol·L-1) was used as solvent media.RESULTS The results showed good linearity(r= 0.9994~0.9997).The average recoveries were 90.4%~102.5% and each RSD was less than 10%.The minimum detection concentrations of chloroform,dichloromethane,1,2-dichloroethane and carbon tetrachloride were 3.0,24.0,7.5 and 0.2 μg·L-1,respectively.CONCLUSION This method is simple,sensitive and accurate.It is suitable for determination of organic solvents and solvent impurities in rhein.