STUDY ON METAL MEMBRANE TEMPERATURE SENSOR AND MICROHEATER FOR PCR CHIP

Yan Weiping,Xue Ning,Shi Xinghui,Liu Junshan,Guo Jihong
DOI: https://doi.org/10.1142/s0218625x08011184
2008-01-01
Surface Review and Letters
Abstract:The development of PCR chip shortens the reaction time of DNA amplification, reduces the bulk of reaction reagents; therefore, it is widely applied to biomedicine and other related fields. Since PCR reaction is completed through three different precise temperature zones periodically, the sensitivity of temperature measurement and the heating property of the metal membrane must be highly regarded. In this paper, three kinds of membranes, Ni , Pt , and Ni – Cr , were fabricated on silicon substrates; their resistance–temperature characteristics and heating properties are discussed. The results show that Ni and Pt films have good linear resistance–temperature relationship; the heating property of Pt film is superior to that of Ni – Cr film. The PCR chip with Pt film microheater can reach 150°C in a short time with 25 V voltage.
What problem does this paper attempt to address?