Design and experimental study of a high/low temperature visualization system in -196 ℃ to 100 ℃

Wang Shanshan,Huang Yonghua,Zhang Liangjun,Wu Jingyi,Xu Yuxiong,Li Suling
DOI: https://doi.org/10.3969/j.issn.1000-6516.2012.03.002
IF: 2.134
2012-01-01
Cryogenics
Abstract:The design and experimental study of a high-low temperature visualization system working in-196 ℃ to +100 ℃ was presented,in which the technologies were integrated,integrating technologies such as vacuum insulation,multilayer insulation and thermal conduction reduction.Under cryogenic environment,heat compensation was provided by a thin film electrical heater and the Joule heat of the camera.For high ambient temperature application,dry nitrogen at room temperature flows through the inner space continuously to cool down the camera.The two modes could be switched consecutively.Illumination was also integrated into the visualization system.Theoretical calculation and experiment show that the system can competent for the desired task with the surface temperature of camera controlled from-30 ℃ to +50 ℃.
What problem does this paper attempt to address?