In-situ measurement of atmospheric hydrofluorocarbon and perfluorocarbons using a custom-made GC-MS system

YAO Bo,ZHOU Lingxi,LI Peichang,XU Lin
IF: 4.236
2012-01-01
Environmental Chemistry
Abstract:A custom-made GC-MS system was designed for in-situ measurement of trace atmospheric hydrofluorocarbons(HFCs) and perfluorocarbons(PFCs).The time resolution of the air samples is 120 min.The precision and accuracy ranges are 0.15%—3.11% and-1.52%—0.63%,respectively.The field measurement results during 2011-02-01—2011-02-28 at the Shangdianzi background station show that the system is able to capture both the background and pollution signals and can be used for in-situ background atmospheric measurements of HFCs and PFCs.
What problem does this paper attempt to address?