Online Monitoring System for Volatile Organic Compounds in the Atmosphere

Xinglong LIU,Limin ZENG,Sihua LU,Xuena YU
DOI: https://doi.org/10.3321/j.issn:0253-2468.2009.12.002
2009-01-01
Abstract:A new online monitoring system for atmospheric volatile organic compounds, including oxygenated volatile organic compounds, was developed with a combination of cryo-trapping-thermal desorption and GC-FID/MS ( Gas chromatography-Flame Ionization Detector/Mass Spectrometry). Water and ozone were removed to avoid interference. Ambient air was collected with an air flow rate of 60 mL. min~(-1) at-15℃ for S minutes, then the sample was desorbed by beating to 110℃ and the volatile organic compounds were detected by GC-FID/MS. The time resolution is 1 h. The system was calibrated with multi-component standards. The calibration results indicated good regression results for target compounds with R~2 = 0. 9137 -0. 9998 and relative standard deviation ( BSD% )below 10%. The newly developed system showed a similar variation of VOC species concentrations to those currently used incommercial online instruments, and the correlation coefficients (r) ranged between 0. 7412 and 0. 9620.
What problem does this paper attempt to address?