On-line Trace Monitoring of Volatile Halogenated Compounds in Air by Improved Thermal Desorption-Gas Chromatography-Mass Spectrometry.

Jiakui Jiang,Yongyan Ji,Xiang Li
DOI: https://doi.org/10.1016/j.chroma.2022.463507
IF: 4.601
2022-01-01
Journal of Chromatography A
Abstract:Volatile halogenated compounds (VHCs) are important industrial chemicals and play a crucial role in potential stratospheric ozone-depletion and global warming. Profiling of VHCs is of great significance but replete with challenges due to the species' richness and diversity. In this study, we developed a novel method employing water removal mode combined with thermal desorption-gas chromatography-mass spectrometry (TD-GC-MS) for on-line measurement of VHCs at the ultratrace level. By removing water with Nafion dryer tandem cold trap device, VHCs could achieve better separation and identification, and detection precision of VHCs lower than 10%. The proposed method exhibited limits of detection for VHCs ranging from 0.1 to 6.2 pptv. Benefiting from the improved trapping efficiency due to water removal, we successfully quantified 34 VHCs at the Shangdianzi background station and achieved a comprehensive assessment of VHCs in ambient air.
What problem does this paper attempt to address?