Sampling-analysis-quality Control Method for Atmospheric Hydrofluorocarbons (Hfcs)

YAO Bo,LI Pei-chang,ZHOU Ling-xi,XU Lin,ZHOU Huai-gang,DONG Fan,ZHOU Li-yan,SHI Qing-feng
DOI: https://doi.org/10.3969/j.issn.1000-6923.2012.09.009
2012-01-01
Abstract:Air samples were flushed and pumped into two-port stainless steel canisters and pressurized to 1.36×105 Pa by a custom-made sampler.The samplers were analyzed by a gas chromatography-mass spectrometry(GC-MS) system in the laboratory.The precisions of 6 major hydrofluorocarbons(HFCs) analysis were 0.24%~1.02%.The concentrations of 6 HFCs in the blank samples were below detection limits of GC-MS system.The influence caused by various inlet pressures fluctuation was corrected.The recoveries for 6 HFCs during sampling-analysis procedure were 99.5%~100.4% and there were not significant changes of the HFCs concentrations during 112-days storage.Air samples from the top of an 80-m tower at Shangdianzi regional station were sampled and analyzed.The concentration differences between two parallel samples were 0.04×10-12~0.16×10-12.The concentrations difference between canister samples and in-situ measurement were-0.17×10-12~-0.86×10-12.The sampling-analysis-quality control method was suitable for high precision atmospheric HFCs observation.
What problem does this paper attempt to address?