Dual channel GC system for measuring background atmospheric CH_4,CO,N_2O and SF_6

FANG Shuangxi,ZHOU Lingxi,ZHANG Fang,YAO Bo,ZHANG Xiaochun,ZANG Kunpeng,XU Lin,LIU Lixin,WEN Min,GU Shuai
DOI: https://doi.org/10.13671/j.hjkxxb.2010.01.007
2010-01-01
Abstract:A high-precision dual channel GC system was set up at Mt. Waliguan Global Atmosphere Watch Station,and used to measure atmospheric CH4,CO,N2O and SF6. The gases were identified by retention time on an Agilent 6890N gas chromatograph. The concentrations of CH4,CO,SF6 were calculated from peak height and N2O from peak area. The RSD (Relative Standard Deviation) of 110 repeated injections were better than 0.04% for CH4 (2000.3×10-9),0.5% for CO (199.8×10-9),0.08% for N2O (320.04×10-9) and 1.80% for SF6 (5.9×10-12). To meet the needs of high precision in-situ measurement,one standard gas injection was applied after every three unknown samples. Results from test operation showed that the system was suitable for background atmospheric CH4,CO and SF6 measurement with high precision,stability and reliability.
What problem does this paper attempt to address?