Evaluation of SO2-, NO3- and NHf measurements using ambient PM2.5 real-time monitoring instruments

袁超,王韬,高晓梅,聂玮,雷国强,徐政,王文兴
IF: 4.236
2012-01-01
Environmental Chemistry
Abstract:Field studies were conducted in Hong Kong to assess two different real-time techniques for measuring SO42- , NOr and NH~ in PM2.5 by comparison with a honeycomb denuder filter-pack system. An ambient ion monitor (AIM, URG9000B ) showed good performance for NO3- and NH4+, but obvious overestimation for SO~-, which was attributed to the positive interference by high ambient SO2 concentrations which could not be totally eliminated in the parallel denuder. To overcome this problem, a series of experiments were conducted, which lead to an improved system including two denuders working in series and with the use of a new denuder liquid (5 mmol" L-1 H202 + 5 mmol-L-1 NaOH). The new system could absorb high concentration of SO2 (even up to 260 p~g-m-3) so that the SO~- measurement results would not be affected. Another monitor for measuring aerosols and gases ( MARGA, Model ADI 2080), developed by ECN, Metrohm and Applikon, was also evaluated. MARGA showed excellent performance for the measurements of SO~- and NI-I~ , but exhibited high NO3- measurement results. The accuracy of NO~- measurements should be evaluated further. This study demonstrated the need for evaluating new measurement techniques in different atmospheric environments to determine the accuracy and precision of the measurements.