An comparative analysis of the accuracy of atmospheric NOx measurements

Wei Cao,Limin Zeng,Yusheng Wu,Min Hu
2013-01-01
Abstract:An intercomparison of atmospheric measurements was conducted using the mainstream instruments of NOx (Model 42i-TL NO-NO2-NOx) and NOy (Model 42i NOy analyzers), both by Thermo Fisher Scientific Inc., at an urban site in Beijing. During the study, concentrations of NO, NOx and NOy were continuously monitored from January to October 2011. Good correlation (r>0.998, p<0.01) of NO measurements produced by the two instruments was obtained, which confirms measurement accuracy of the two devices. By comparing the NOy measurement results obtained under the conditions of attaching and removing a particulate filter from the sample inlet of the Model 42i NOy analyzer with that of the NOx measurements using the NO-NO2-NOx analyzer, the following correlation results are obtained: [NOy]=0.989×[NOx], R2=0.993 and [NOy]=1.134×[NOx], R2=0.959. Two conclusions are drawn from this study: (1) The difference between NOx and NOy is caused by particulate nitrate which accounts for around 10 percent of the total NOx.(2) Concentration of NOx measured using the current Model 42i-TL analyzer is higher than its real values, and is closer to concentration of NOy. Using the correction method of subtracting concentrations of the main NOy components (e.g. HONO, HNO3, PAN, PPN) from that of NO2 (based on data collected during August 6-15, 2011), it could be concluded that concentration of NO2 in summer are overestimated by 7 percent (R2=0.968).