Simultaneous On-Line FTIR Measurements of Multicomponent Pollutants in Wet Gas Mixtures

李天津,禚玉群,陈昌和,徐旭常
DOI: https://doi.org/10.3321/j.issn:1000-0054.2008.11.024
2008-01-01
Abstract:An on-line quantitative analysis method was developed using FTIR to simultaneously measure NO, SO2, NH3, NO2 and N2O in wet gas mixtures in the simultaneous SO2 and NOx control experiments. The calibration range of each component covers the typical concentration ranges commonly observed in laboratory-scale flue gas desulfurization and denitrification analyses. Experimental results show that accurate concentrations of NO, SO2, NH3, NO2 and N2O in wet gas mixtures can be simultaneously measured on-line by appropriately selecting the sampling parameters and the parameters of this method. The accuracy of the single gas component measurement in the absence of H2O is estimated to be ±2%, with the NO measurement accuracy not reduced in gas mixtures with 5% H2O. The simultaneous measurement accuracies of NO, SO2, and NH3 are ±2%, ±3%, and ±3%, respectively, in the presence of 5% H2O and 14% CO2.
What problem does this paper attempt to address?