Simultaneous Online Measurement of H 2 O and CO 2 in the Humid CO 2 Adsorption/Desorption Process

Yu Qingni,Ye Sha,Zhu Jingke,Lei Lecheng,Yang Bin
DOI: https://doi.org/10.2116/analsci.31.757
IF: 1.967
2015-01-01
Analytical Sciences
Abstract:A dew point meter (DP) and an infrared (IR) CO 2 analyzer were assembled in a humid CO 2 adsorption/desorption system in series for simultaneous online measurements of H 2 O and CO 2 , respectively. The humidifier, by using surface-flushing on a saturated brine solution was self-made for the generation of humid air flow. It was found that by this method it became relatively easy to obtain a low H 2 O content in air flow and that its fluctuation could be reduced compared to the bubbling method. Water calibration for the DP-IR detector is necessary to be conducted for minimizing the measurement error of H 2 O. It demonstrated that the relative error (RA) for simultaneous online measurements H 2 O and CO 2 in the desorption process is lower than 0.1%. The high RA in the adsorption of H 2 O is attributed to H 2 O adsorption on the transfer pipe and amplification of the measurement error. The high accuracy of simultaneous online measurements of H 2 O and CO 2 is promising for investigating their co-adsorption/desorption behaviors, especially for direct CO 2 capture from ambient air.